Title: 運動負荷時における血中乳酸濃度の乳酸迅速測定器(LACTATE PRO)によるモニタリング(自然科学系)
Other Titles: Monitoring of Blood Lactate Concentration during Physical Exercise with the Device "LACTATE PRO"
Authors: 久原, 巻彦
佐藤, 友治
小西, 利佳
杉原, 香文
前河, 裕一
小林, 惠美
奥田, 清
Author's alias: Kuhara, Makihiko
Sato, Yuji
Konishi, Rika
Sugihara, Kofumi
Maekawa, Yuichi
Kobayashi, Emi
Okuda, Kiyoshi
Keywords: anaerobic threshold
lactate threshold
electrode strip method
Issue Date: 31-Dec-1997
Publisher: 大阪府立看護大学医療技術短期大学部
Citation: 大阪府立看護大学医療技術短期大学部紀要. 1997, 3, p.51-56
Abstract: We tested the device "LACTATE PRO" for rapid estimation of the lactate threshold in blood. Blood samples were obtained from eight healthy volunteers at various times during and after a treadmill exercise. Applying these samples to the LACTATE PRO measurement immediately after blood collection, each lactate level in the blood at the time of blood collection could be known just I min after the blood collection. Lactate levels measured with LACTATE PRO were reproducible. A good correlation was found between the lactate levels obtained by the LACTATE PRO method and those obtained by the standard enzymatic method. Assay for lactate in various samples with LACTATE PRO revealed that there was not a significant correlation between the lactate levels in venous blood and those in capillary blood.